Key features of IC Plus detectors

Available with four electrode lengths
Adjustable electrodes separations
Easily replaceable kapton windows.
Operating pressure range :
Accept any gas filling provided it is
........compatible with the aluminum body.
Operating voltage up to 1700 volts.
Guard ring on signal electrode.
Can be used as stand alone or with the piggy
....... back electronics module.
: 10, 50, 150 and 300 mm.
: 10, 14 and 18 mm.

: 0.7 to 1.3 bars.

Electronics module

It minimizes noise due to ground loops. It includes a preamplifier, a voltage to frequency converter, a high voltage
... supply and the RS232 communication port for complete remote control from the host computer.
It is fully compatible with X2000 detection electronics and can be chained on the same RS232 line.

Current range
Output signal (voltage)
Input bias current
Long term offset stability
Input current noise F = 0.1 to 10 Hz
Input impedance
Input capacitance
Output impedance
Voltage to frequency converter
........output frequency
High voltage supply
RS232 line
Address chamber
Power requirements
: from 0.1 pA to 0.1mA in six overlapping ranges
: 1 mV to 10 V
: 100 fA at 25°C typical
: 15 µV / month
: 10 fA peak to peak on highest gain range
: 1012 ohms
: 1 pF
: 50 ohms short circuit protected
: 0 to 1 MHz

: adjustable from 0 to 1700 volts, short circuit protected
: adjustment and read back of the electronics settings
: can be selected from 0 to 15
: 24 VDC - 150 mA from a small power supply provided.
110/220 V - 50/60 Hz

__________________________________________________________________________________________________________

__________________________________________________________________________________________________________

50 mm and 300 mm ionisation chambers. Electronics module
Picture shown :

50 mm and 300 mm ionisation chambers
Electronics module

These ionisation chambers are especially designed for intensity monitoring applications.
The IC plus range offers premium performance at reasonable cost.
As shown on the picture, front end electronics can be directly plugged on chambers.
These instruments are promoted and sold worldwide by FMB Oxford.

  Javascript DHTML Drop Down Menu Powered by dhtml-menu-builder.com
Headquarters
Home
Company
Quality
Contact

 

 

 

 

 

 

 

 

 

 

Sapphire, sapphire crystal, sapphire crystal growth equipment, sapphire crystal furnace, radiation detector, ionisation chamber, ic, ionization chamber, scintillation detector, compton suppression, anti compton, x ray detector, gamma ray detector, cosmic suppression, crystal scintillation detector, detector using liquid scintillator, detector using plastic scintillator, scintillator, floating zone, gas blower, cyberstar, ionisation chambers, anti-compton detectors, hot air generator, gas tube,scintillation, radiation detection